Response of a tissue equivalent proportional counter to neutrons. Academic Article uri icon

abstract

  • The absorbed dose as a function of lineal energy was measured at the CERN-EC Reference-field Facility (CERF) using a 512-channel tissue equivalent proportional counter (TEPC), and neutron dose equivalent response evaluated. Although there are some differences, the measured dose equivalent is in agreement with that measured by the 16-channel HANDI tissue equivalent counter. Comparison of TEPC measurements with those made by a silicon solid-state detector for low linear energy transfer particles produced by the same beam, is presented. The measurements show that about 4% of dose equivalent is delivered by particles heavier than protons generated in the conducting tissue equivalent plastic.

published proceedings

  • Radiat Meas

altmetric score

  • 3

author list (cited authors)

  • Badhwar, G. D., Robbins, D. E., Gibbons, F., & Braby, L. A.

citation count

  • 11

complete list of authors

  • Badhwar, GD||Robbins, DE||Gibbons, F||Braby, LA

publication date

  • December 2002