An Insertable P-31 RF coil for Dual-frequency Magnetic Resonance Imaging & Spectroscopy Conference Paper uri icon

abstract

  • Application of nuclear magnetic resonance spectroscopy in metabolic studies usually includes simultaneous excitation at two different frequencies. RF coupling between the two coils has been a problem in the past, resulting in reduced field strength from the1H coil. A modular surface coil was designed and constructed for1H and31P at 4.7 Tesla; RF traps were designed and optimized to minimize the field-degrading coupling current, and trap performance was analyzed in terms of improvement in signal-to-noise-ratio and field sensitivity pattern of the proton coil. The modular design in conjunction with the optimized traps permits the possibility of insertion of the phosphorus coil as an "add on" into any other hydrogen coil. 2008 IEEE.

name of conference

  • 2008 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society

published proceedings

  • 2008 30TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-8

author list (cited authors)

  • Dabirzadeh, A., Chang, C., & McDougall, M. P.

citation count

  • 6

complete list of authors

  • Dabirzadeh, Arash||Chang, Chieh-Wei||McDougall, Mary P

publication date

  • December 2008