Efficient method to design RF pulses for parallel excitation MRI using gridding and conjugate gradient. uri icon

abstract

  • Parallel excitation (pTx) techniques with multiple transmit channels have been widely used in high field MRI imaging to shorten the RF pulse duration and/or reduce the specific absorption rate (SAR). However, the efficiency of pulse design still needs substantial improvement for practical real-time applications. In this paper, we present a detailed description of a fast pulse design method with Fourier domain gridding and a conjugate gradient method. Simulation results of the proposed method show that the proposed method can design pTx pulses at an efficiency 10 times higher than that of the conventional conjugate-gradient based method, without reducing the accuracy of the desirable excitation patterns.

published proceedings

  • Quant Imaging Med Surg

author list (cited authors)

  • Feng, S., & Ji, J.

complete list of authors

  • Feng, Shuo||Ji, Jim