Water vapor (2) H, (18) O and (17) O measurements using an off-axis integrated cavity output spectrometer - sensitivity to water vapor concentration, delta value and averaging-time. Academic Article uri icon

abstract

  • RATIONALE: High-precision analysis of atmospheric water vapor isotope compositions, especially (17) O values, can be used to improve our understanding of multiple hydrological and meteorological processes (e.g., differentiate equilibrium or kinetic fractionation). This study focused on assessing, for the first time, how the accuracy and precision of vapor (17) O laser spectroscopy measurements depend on vapor concentration, delta range, and averaging-time. METHODS: A Triple Water Vapor Isotope Analyzer (T-WVIA) was used to evaluate the accuracy and precision of (2) H, (18) O and (17) O measurements. The sensitivity of accuracy and precision to water vapor concentration was evaluated using two international standards (GISP and SLAP2). The sensitivity of precision to delta value was evaluated using four working standards spanning a large delta range. The sensitivity of precision to averaging-time was assessed by measuring one standard continuously for 24hours. RESULTS: Overall, the accuracy and precision of the (2) H, (18) O and (17) O measurements were high. Across all vapor concentrations, the accuracy of (2) H, (18) O and (17) O observations ranged from 0.10 to 1.84, 0.08 to 0.86 and 0.06 to 0.62, respectively, and the precision ranged from 0.099 to 0.430, 0.009 to 0.080 and 0.022 to 0.054, respectively. The accuracy and precision of all isotope measurements were sensitive to concentration, with the higher accuracy and precision generally observed under moderate vapor concentrations (i.e., 10000-15000ppm) for all isotopes. The precision was also sensitive to the range of delta values, although the effect was not as large compared with the sensitivity to concentration. The precision was much less sensitive to averaging-time than the concentration and delta range effects. CONCLUSIONS: The accuracy and precision performance of the T-WVIA depend on concentration but depend less on the delta value and averaging-time. The instrument can simultaneously and continuously measure (2) H, (18) O and (17) O values in water vapor, opening a new window to better understand ecological, hydrological and meteorological processes. Copyright 2016 John Wiley & Sons, Ltd.

published proceedings

  • Rapid Commun Mass Spectrom

altmetric score

  • 1.25

author list (cited authors)

  • Tian, C., Wang, L., & Novick, K. A.

citation count

  • 20

complete list of authors

  • Tian, Chao||Wang, Lixin||Novick, Kimberly A

publication date

  • October 2016

publisher