Generalized shift-rotation absolute measurement method for high-numerical-aperture spherical surfaces with global optimized wavefront reconstruction algorithm. Academic Article uri icon

abstract

  • In this paper, a generalized shift-rotation absolute measurement method is proposed to measure the absolute surface shape of high-numerical-aperture spherical surfaces. Based on the wavefront difference method, the high order misalignment aberrations can be removed from the measurements. Our generalized shift-rotation absolute measurement process only needs one rotational measurement position and one translational measurement position. A wavefront reconstruction method based on the self-adaptive differential evolution algorithm is proposed to calculate the Zernike polynomials coefficient ai of the absolute surface shape Wtest(x,y), the rotation angle , the translation x along the x axis, and the translation y along the y axis. The translation error and rotation error in other absolute measurement methods are avoided using our generalized shift-rotation absolute measurement method. Experimental absolute results of the test surface and reference surface are given and the difference of reference surface shapes between two testings in experiments is 0.12 nm root mean square.

published proceedings

  • Opt Express

author list (cited authors)

  • Yang, Z., Du, J., Tian, C., Dou, J., Yuan, Q., & Gao, Z.

citation count

  • 12

complete list of authors

  • Yang, Zhongming||Du, Jinyu||Tian, Chao||Dou, Jiantai||Yuan, Qun||Gao, Zhishan

publication date

  • October 2017