Array biosensor for detection of biohazards. Academic Article uri icon

abstract

  • A fluorescence-based biosensor has been developed for simultaneous analysis of multiple samples for multiple biohazardous agents. A patterned array of antibodies immobilized on the surface of a planar waveguide is used to capture antigen present in samples; bound analyte is then quantified by means of fluorescent tracer antibodies. Upon excitation of the fluorophore by a small diode laser, a CCD camera detects the pattern of fluorescent antibody:antigen complexes on the waveguide surface. Image analysis software correlates the position of fluorescent signals with the identity of the analyte. This array biosensor has been used to detect toxins, toxoids, and killed or non-pathogenic (vaccine) strains of pathogenic bacteria. Limits of detection in the mid-ng/ml range (toxins and toxoids) and in the 10(3)-10(6) cfu/ml range (bacterial analytes) were achieved with a facile 14-min off-line assay. In addition, a fluidics and imaging system has been developed which allows automated detection of staphylococcal enterotoxin B (SEB) in the low ng/ml range.

published proceedings

  • Biosens Bioelectron

altmetric score

  • 3

author list (cited authors)

  • Rowe-Taitt, C. A., Golden, J. P., Feldstein, M. J., Cras, J. J., Hoffman, K. E., & Ligler, F. S.

citation count

  • 149

complete list of authors

  • Rowe-Taitt, CA||Golden, JP||Feldstein, MJ||Cras, JJ||Hoffman, KE||Ligler, FS

publication date

  • January 2000