Variation of fluorescence in tissue with temperature.
Academic Article
Overview
Research
Identity
Additional Document Info
Other
View All
Overview
abstract
BACKGROUND AND OBJECTIVE: Previous studies demonstrated a decrease in fluorescence intensity as tissue temperature increased. In vitro samples were increased from room temperature and in vivo canine liver from body temperature. This study investigated variations in fluorescence intensity with temperatures starting at 14C and compared in vivo and in vitro results for consistency. STUDY DESIGN/MATERIAL AND METHODS: A fiber optic-based noninvasive system was used to characterize the temperature effect on tissue fluorescence in hamster dorsal skin in vivo, and in sclera and cornea of enucleated pig eyes in vitro. As tissue was allowed to progress through the temperature range of 14-42C, the spectra of auto-fluorescence with respect to temperature was sampled every 1-2 minutes. A pulsed nitrogen laser was used to excite fluorescence through a fiber optic probe with a source-detector aperture separation of 370m. RESULTS: Fluorescence intensity decreased as temperature increased from 14 to 42C in a phantom containing Rhodamine B dye. Results from both in vivo and in vitro tissue followed the same trend of decreasing intensity as tissue temperature increased from 14C. Spectral intensity lineshape changed around 450nm due to absorption from tissue. CONCLUSION: Cooling a tissue increased fluorescence intensity of skin in vivo, in all experiments. In vitro results were consistent with in vivo measurements.