Myocardial suppression in vivo by spin locking with composite pulses. Academic Article uri icon

abstract

  • Improved myocardium-blood contrast by myocardial suppression resulting from T1 rho-weighting in contrast-enhanced, gradient-echo, bright-blood cine images, acquired at 1.5T, is shown. In the standard images, blood has twice the intensity of muscle. In similar T1 rho-weighted images, it has 3-4 times the intensity of muscle. A composite spin-lock pulse before each observation pulse provides T1 rho-weighting. A typical pulse was: 90y-135x-360x-135x-90(-y) with element durations: 0.84, 1.26, 8.12, 1.26, and 0.84 ms. The tolerance of this composite pulse to shimmering and frequency errors allows spin locking with comparatively weak RF and therefore low specific absorption rate (SAR). Initial clinical evaluation on patients with poor ventricular function demonstrates both a qualitative and quantitative improvement in delineation of myocardial borders.

published proceedings

  • Magn Reson Med

altmetric score

  • 3

author list (cited authors)

  • Dixon, W. T., Oshinski, J. N., Trudeau, J. D., Arnold, B. C., & Pettigrew, R. I.

citation count

  • 75

complete list of authors

  • Dixon, WT||Oshinski, JN||Trudeau, JD||Arnold, BC||Pettigrew, RI

publication date

  • July 1996