Effects of acceleration on the accuracy of MR phase velocity measurements. Academic Article uri icon

abstract

  • Acceleration in blood flow can affect the accuracy of phase velocity measurements. Convective acceleration is due to changes in flow geometry and is independent of the time-varying acceleration caused by flow pulsatility. To analyze the effects of convective acceleration on flow velocity measurements, phase velocity measurements were obtained in steady laminar flow in the convergent segment of a 90%, hourglass-shaped stenosis phantom at a Reynolds number of 1,500. Measurements at the stenosis indicated that convective acceleration caused the measured values of average cross-sectional velocity to deviate as much as 37% from the theoretical values. The magnitude of the error could be accounted for by including the convective acceleration term in the phase shift equation. Convective acceleration effects should not be ignored in flow velocity measurements through stenoses, even when time-dependent acceleration due to flow pulsatility can be neglected.

published proceedings

  • J Magn Reson Imaging

author list (cited authors)

  • Oshinski, J. N., Ku, D. N., Bohning, D. E., & Pettigrew, R. I.

citation count

  • 53

complete list of authors

  • Oshinski, JN||Ku, DN||Bohning, DE||Pettigrew, RI

publication date

  • November 1992