A method for intensity calibration of an electron spectrometer with multi-angle detection Academic Article uri icon

abstract

  • 2015 Elsevier B.V. All rights reserved. A special electrostatic electron spectrometer designed for precise and unique experiments and an intensity calibration method for universal application in electron spectroscopy are presented. The upgrade of the analyzer enables the intensity calibration at arbitrary electron energies using elastically scattered electrons. In order to test the calibration procedure the double differential (in energy and angle) ionization cross sections (DDCS) of electrons ejected from 300 keV proton-argon collisions were measured and compared with the data of Rudd et al. [1]. The good agreement between the two data sets verifies the applicability of the calibration method.

published proceedings

  • Nuclear Instruments and Methods in Physics Research Section B Beam Interactions with Materials and Atoms

author list (cited authors)

  • brk, L., Buhr, T., Kvr, . ., Balog, R., Hatvani, D., Herczku, P., Kovcs, S., & Ricz, S.

citation count

  • 2

complete list of authors

  • Ábrók, Levente||Buhr, Ticia||Kövér, Ákos||Balog, Róbert||Hatvani, Dávid||Herczku, Péter||Kovács, Sándor||Ricz, Sándor

publication date

  • February 2016