Development and validation of a high-performance liquid chromatography method for the quantification of talazoparib in rat plasma: Application to plasma protein binding studies.
Overview
Research
Identity
Additional Document Info
Other
View All
Overview
abstract
A sensitive and selective RP-HPLC method has been developed and validated for the quantification of a highly potent poly ADP ribose polymerase inhibitor talazoparib (TZP) in rat plasma. Chromatographic separation was performed with isocratic elution method. Absorbance for TZP was measured with a UV detector (SPD-20A UV-vis) at a max of 227 nm. Protein precipitation was used to extract the drug from plasma samples using methanol-acetonitrile (65:35) as the precipitating solvent. The method proved to be sensitive and reproducible over a 100-2000 ng/mL linearity range with a lower limit of quantification (LLQC) of 100 ng/mL. TZP recovery was found to be >85%. Following analytical method development and validation, it was successfully employed to determine the plasma protein binding of TZP. TZP has a high level of protein binding in rat plasma (95.76 0.38%) as determined by dialysis method.