Aerosol reduction during air polishing. Academic Article uri icon

abstract

  • OBJECTIVES: Concern has been expressed about the obvious aerosol cloud that is generated during air polishing of teeth. These concerns have included infection control, potential systemic effects, and environmental surface contamination associated with the particles of sodium bicarbonate powder, water, and patient-generated material. This in vitro study evaluated an aerosol reduction device designed for use during air polishing. METHOD AND MATERIALS: Eight artificial teeth on a Dentoform model were polished with and without the aerosol reduction device. The polishing was performed within a plastic enclosure that had a grid on the sides. The water supply for the air polisher contained a 1% fluorescein solution. The aerosol contamination was determined by counting the number of squares that showed fluorescent spots under an ultraviolet light. RESULTS: The mean contamination without the aerosol reduction device was 175.59 cm2. The mean contamination with the aerosol reduction device was 4.37 cm2, a greater than 97% reduction in aerosol contamination. CONCLUSION: The aerosol reduction device significantly decreased the contamination produced during air polishing.

published proceedings

  • Quintessence Int

author list (cited authors)

  • Harrel, S. K., Barnes, J. B., & Rivera-Hidalgo, F

citation count

  • 10

complete list of authors

  • Harrel, SK||Barnes, JB||Rivera-Hidalgo, F

publication date

  • September 1999