Sub-femtomole peptide detection in ion mobility-time-of-flight mass spectrometry measurements.
Academic Article
Overview
Research
Identity
Additional Document Info
View All
Overview
abstract
Sensitivity and detection limit enhancements are obtained for peptides by performing high repetition rate (150 Hz) matrix-assisted laser desorption/ionization (MALDI) coupled with ion mobility-time-of-flight mass spectrometry. Absolute limits of detection (3sigma) for model peptides are on the order of 0.1 fmol of peptide deposited and represent a factor of 40-60 improvement over data obtained using typical low repetition (20 Hz) MALDI. This increase in sensitivity is demonstrated for two-dimensional MALDI-IM-TOFMS peptide mass mapping of bovine hemoglobin.