Ambient noise interferes with auscultatory blood pressure measurement during exercise.
Academic Article
Overview
Research
Identity
Additional Document Info
Other
View All
Overview
abstract
This study was designed to investigate whether the acoustical characteristics of the Korotkoff sounds (K-sounds) were altered during exercise and/or masked by the ambient noise. After signing informed consent, 11 subjects (8 females, 3 males; 27 +/- 2 yr; 166.2 +/- 3.2 cm; 62 +/- 5 kg; means +/- SD) underwent a cycle ergometer exercise test that increased in workload by 30 W every 3 min until volitional fatigue. Heart rate, auscultatory systolic (SBP) and diastolic blood pressure (DBP), and oxygen consumption were monitored 1 and 2 min into each work stage. The auscultatory K-sounds were recorded with a microphone mounted in a stethoscope tube for later frequency (Hz) and sound pressure level (dB SPL) analysis. Frequency and SPL of ambient noise (99 +/- 13 Hz and 64 +/- 1 db at maximum, respectively) increased during the exercise test to magnitudes similar to the SBP and DBP K-sounds (166 Hz, 66 db; and 128 Hz, 69 db, respectively). Additionally, the ambient noise was responsible for a significant damping of the frequency and SPL of the measured blood pressure K-sounds and a rise in the measured frequency of the SBP K-sounds. Furthermore, we observed "inaudible" K-sounds at lower frequencies than adjoining audible K-sounds (100 Hz vs 126 Hz), supporting the known underestimation of SBP by auscultation. The increase in ambient noise during exercise testing dampens and may mask the auscultatory K-sounds, thus making detection of the proper K-sounds during exercise difficult at best. Furthermore, the presence of inaudible K-sounds may further explain the published discrepancies between auscultatory and intraarterial blood pressure measurements during exercise.