Real-time, closed-loop dual-wavelength optical polarimetry for glucose monitoring. Academic Article uri icon

abstract

  • The development of a real-time, dual-wavelength optical polarimetric system to ultimately probe the aqueous humor glucose concentrations as a means of noninvasive diabetic glucose monitoring is the long-term goal of this research. The key impact of the work is the development of an approach for the reduction of the time-variant corneal birefringence due to motion artifact, which is still a limiting factor preventing the realization of such a device. Our dual-wavelength approach utilizes real-time, closed-loop feedback that employs a classical three-term feedback controller and efficiently reduces the effect of motion artifact that appears as a common noise source for both wavelengths. In vitro results are shown for the open-loop system, and although the dual-wavelength system helps to reduce the noise, it is shown that closed-loop control is necessary to bring the noise down to a sufficient level for physiological monitoring. Specifically, in vitro measurement results with the closed-loop dual-wavelength approach demonstrate a sensitivity of 12.8 mg/dl across the physiologic glucose range in the presence of time-variant test cell birefringence. Overall, it is shown that this polarimetric system has the potential to be used as a noninvasive measure of glucose for diabetes.

published proceedings

  • J Biomed Opt

altmetric score

  • 3

author list (cited authors)

  • Malik, B. H., & Cot, G. L.

citation count

  • 56

complete list of authors

  • Malik, Bilal H||Coté, Gerard L

publication date

  • December 2010