An insertable nonlinear gradient coil for phase compensation in SEA imaging. Academic Article uri icon

abstract

  • In magnetic resonance imaging with array coils with many elements, as the radiofrequency (RF) coil dimensions approach the voxel dimensions, the phase gradient due to the magnetic field pattern of the coil causes signal cancellation within each voxel. In single echo acquisition (SEA) imaging with coil arrays, a gradient pulse can be applied to compensate for this effect. However, because RF coil phase varies with distance from the array and reverses on opposite sides of a dual-sided array, this method of phase compensation can be optimized for only a single slice at a time. In this study, a nonlinear gradient coil was implemented to provide spatially varying phase compensation to offset the coil phase with slice position for dual-sided arrays of narrow coils. This nonlinear gradient coil allows the use of one phase compensation pulse for imaging multiple slices through a slab, and, importantly, is shown to enable simultaneous SEA imaging from opposite sides of a sample using a dual-sided receive array.

published proceedings

  • IEEE Trans Biomed Eng

author list (cited authors)

  • Bosshard, J. C., McDougall, M. P., & Wright, S. M.

citation count

  • 1

complete list of authors

  • Bosshard, John C||McDougall, Mary Preston||Wright, Steven M

publication date

  • January 2014