A mass-selective variable-temperature drift tube ion mobility-mass spectrometer for temperature dependent ion mobility studies. Academic Article uri icon

abstract

  • A hybrid ion mobility-mass spectrometer (IM-MS) incorporating a variable-temperature (80-400K) drift tube is presented. The instrument utilizes an electron ionization (EI) source for fundamental small molecule studies. Ions are transferred to the IM-MS analyzer stages through a quadrupole, which can operate in either broad transmission or mass-selective mode. Ion beam modulation for the ion mobility experiment is accomplished by an electronic shutter gate. The variable-temperature ion mobility spectrometer consists of a 30.2cm uniform field drift tube enclosed within a thermal envelope. Subambient temperatures down to 80K are achievable through cryogenic cooling with liquid nitrogen, while elevated temperatures can be accessed through resistive heating of the envelope. Mobility separated ions are mass analyzed by an orthogonal time-of-flight (TOF) mass spectrometer. This report describes the technological considerations for operating the instrument at variable temperature, and preliminary results are presented for IM-MS analysis of several small mass ions. Specifically, mobility separations of benzene fragment ions generated by EI are used to illustrate significantly improved (greater than 50%) ion mobility resolution at low temperatures resulting from decreased diffusional broadening. Preliminary results on the separation of long-lived electronic states of Ti(+) formed by EI of TiCl(4) and hydration reactions of Ti(+) with residual water are presented.

published proceedings

  • J Am Soc Mass Spectrom

author list (cited authors)

  • May, J. C., & Russell, D. H.

citation count

  • 59

complete list of authors

  • May, Jody C||Russell, David H

publication date

  • July 2011