A high repetition rate (1 kHz) microcrystal laser for high throughput atmospheric pressure MALDI-quadrupole-time-of-flight mass spectrometry.
Academic Article
Overview
Research
Identity
Additional Document Info
Other
View All
Overview
abstract
Sample throughput has been increased in many areas of proteomics, but the last significant advance in lasers used for matrix-assisted laser desorption/ionization (MALDI) was the introduction of cartridge-type N2 lasers (337 nm, 4-ns pulse widths, 1-30-Hz repetition rates) more than a decade ago. This report describes the application of a 1-kHz repetition rate Nd:YAG laser (355 nm, <500-ps pulse widths) for atmospheric pressure MALDI-QqTOFMS, and data obtained are compared to a conventional nitrogen laser. For example, the signal intensity for angiotensin II using the 1-kHz laser was in some cases enhanced by a factor of 80 and high-quality data could be obtained in as little as 1 s.